Cold Box Passes Comprehensive 3rd Party IOQ and PQ Validation Protocol for Use in Pharmaceutical and Life Science Industries

Cold Box Express, LLC recently completed and passed a comprehensive 3rd party Installation Operational Qualification and Performance Qualification or IOQ and PQ Validation Protocol effort conducted in a climate controlled setting. In summary, the tests ran for 60 hours with a payload temperature set point of 5⁰C. With 22 Sensitech TempTales (TT4) distributed along the payload compartment, these sensors captured data every 5 minutes. After 60 hours under these simulated real-world conditions, the Cold Box passed all IOQ and PQ Validation Protocol tests by maintaining the temperature set point within the auditable tolerance range. A thermal decay evaluation of our Cold Box shipper followed for an additional 12 hours.
Click here for more information.
Click here to read the protocol.